RX01/LX01 SERIES RF POWER SUPPLY OPERATOR’S MANUAL
ANALOG INTERFACE CONNECTOR PIN LIST
PIN SIGNAL NAME DESCRIPTION
Measure monitor voltage with respect to pin 23 (REFRET).
12 FEEDBACK External feedback voltage signal. Analog input, 0 to +10.0VDC .
Use pin 16 or 17 or (GNDI) for return reference.
The RF Power Supply will automatically adjust its output power
to maintain the FEEDBACK signal’s magnitude at the same
level as the SETPOINT signal magnitude. The PROBE voltage
will be displayed on the front panel. The PROBE attenuation
factor can be configured from the front panel.
The external feedback signal is derived from a voltage probe
(RF or DC) located elsewhere in the plasma or process system.
Refer to the controls section for detailed instructions on how to
configure and use this mode.
Note: The feedback voltage polarity must match the setpoint
input’s (pin 13) polarity for proper operation of voltage control
mode. The RF Power Supply’s external feedback circuitry is not
designed to function with Negative (-) polarity signals applied to
this pin.
13 SETPOINT Power or Voltage setpoint input. Analog, high-impedance,
differential input with selectable 0 to +5.0VDC or 0 to +10.0VDC
range via front panel controls.
Refer to the controls section of the operator’s manual for
detailed instructions on how to configure and use this mode
NOTE: SETRET (pin 25) MUST
be referenced to common or
ground at the setpoint voltage source (system controller) or the
RF output power will behave erratically.
Refer to the Analog Setpoint Sensitivity subsection of the
appropriate model specification for setpoint sensitivity details.
NOTE: Feedback voltage range and polarity must match
setpoint voltage range and polarity for proper operation in
voltage control mode.
Pin 13 is the positive (+) input of the differential setpoint
amplifier.
Active only in ANALOG control mode.
14 MAINS INTLK 2
R101, L101, R301,
L301 only
A contact closure between pin 1 and pin 14 is required to allow
AC mains power to engage. An open circuit between pin 1 and
pin 14 disables AC mains power. 24VAC current loop, 100 mA
maximum current
This signal is active in Panel, Analog, or Serial control modes
NOTE:
Applicable to models R101, L101, R301, L301 only.
No Connection
R101MKII,
L101MKII,
No Connection
Page 90
Seren IPS Inc.
6100220000 Rev. 1.09