RX01/LX01 SERIES RF POWER SUPPLY OPERATOR’S MANUAL
R2001H Specifications
Coolant Water Fittings: Type 3/8” N.P.T. Female threaded pipe
External interlock provided for disabling RF
output power
Interface
Connections:
Power
Monitor
Scaling:
Reflected Power Monitor
Coolant Water Supply
Line Inside Diameter:
3/8” [9.52mm] Inside Diameter (I.D.) or larger.
Waste Heat into Coolant
Water
1600 Watts (91 BTU/minute) maximum
Note: The waste heat does not peak at 2000 Watts
of output power - the waste heat peaks at 50% to
80% of full power or under certain reflected
power conditions.
Interlocks: External:
Valve (optional): Prevents/disables RF Power output if external
water valve is missing or un plugged. (standard
units ignore this interlock – consult factory if
valve interlock is required)
Analog Interface: 25 pin Female D-Sub Connector
Serial Interface: 9 pin Female D-Sub Connector
CEX Input: Female type BNC connector, 3 to 8V Peak-to-
Peak sine wave, 13.56MHz or 1.7-2.1MHz, 50
Ohms Nominal Impedance
CEX Output: Female type BNC connector, 5 to 8V Peak-to-
Peak, 13.56MHz or 1.7-2.1MHz into 50 Ohms
VALVE 4-Pin Male Disk-Drive Connector
Forward Power Monitor Monitor output is linearly proportional to 0 to
2000 Watts of forward power.
Output scaling is 5.0VDC at 2000 Watts
(2.5mV/Watt) or 10.0VDC at 2000 Watts
(5.0mV/Watt), depending on the analog voltage
range selected. (Default scaling)
Monitor output is linearly proportional to 0 to 500
Watts of reflected power.
Output scaling is 5.0VDC at 500 Watts
(10mV/Watt) or 10.0VDC at 500 Watts
(20mV/Watt), depending on the analog voltage
range selected. (Default scaling)
Page 155
Seren IPS Inc.
6100220000 Rev. 1.09